Description

An issue was discovered on Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices. When they are powered on, an X11 display server is started. The display server listens on all network interfaces and is accessible over port 6000. The X11 access control list, by default, allows connections from 127.0.0.1 and 192.168.2.15. If a device is powered on and later connected to a network with DHCP, the device may not be assigned the 192.168.2.15 IP address, leaving the display server accessible by other devices on the network. The exposed X11 display server can then be used to gain root privileges and the ability to execute code remotely by interacting with matchbox-desktop and spawning a terminal. NOTE: This vulnerability only affects products that are no longer supported by the maintainer.

INFO

Published Date :

2025-12-04T00:00:00.000Z

Last Modified :

2025-12-05T20:03:48.993Z

Source :

mitre
AFFECTED PRODUCTS

The following products are affected by CVE-2025-54304 vulnerability.

Vendors Products
Thermofisher
  • Ion Torrent Onetouch 2
  • Ion Torrent Onetouch 2 Firmware

CVSS Vulnerability Scoring System

Detailed values of each vector for above chart.
Attack Vector
Attack Complexity
Privileges Required
User Interaction
Scope
Confidentiality Impact
Integrity Impact
Availability Impact