Description

An issue was discovered on Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices. They run an SSH server accessible over the default port 22. The root account has a weak default password of ionadmin, and a password change policy for the root account is not enforced. Thus, an attacker with network connectivity can achieve root code execution. NOTE: This vulnerability only affects products that are no longer supported by the maintainer.

INFO

Published Date :

2025-12-04T00:00:00.000Z

Last Modified :

2025-12-05T20:06:53.549Z

Source :

mitre
AFFECTED PRODUCTS

The following products are affected by CVE-2025-53963 vulnerability.

Vendors Products
Thermofisher
  • Ion Torrent Onetouch 2
  • Ion Torrent Onetouch 2 Firmware

CVSS Vulnerability Scoring System

Detailed values of each vector for above chart.
Attack Vector
Attack Complexity
Privileges Required
User Interaction
Scope
Confidentiality Impact
Integrity Impact
Availability Impact