Description
A "Privilege boundary violation" vulnerability is identified affecting multiple Radiometer Products. Exploitation of this vulnerability gives a user with physical access to the analyzer, the possibility to gain unauthorized access to functionalities outside the restricted environment. The vulnerability is due to weakness in the design of access control implementation in application software. Other related CVE's are CVE-2025-14096 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required configuration for Exposure: Physical access to the analyzer is needed. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided working proof-of-concept. Radiometer is not aware of any publicly available exploit at the time of publication. Note: CVSS score 6.8 when underlying OS is Windows 7 or Windows XP Operating systems and CVSS score 5.7 when underlying OS is Windows 8 or Windows 10 operating systems.
INFO
Published Date :
2025-12-17T11:45:43.341Z
Last Modified :
2025-12-17T21:46:57.088Z
Source :
Radiometer
AFFECTED PRODUCTS
The following products are affected by CVE-2025-14095 vulnerability.
| Vendors | Products |
|---|---|
| Microsoft |
|
| Radiometer |
|
REFERENCES
Here, you will find a curated list of external links that provide in-depth information to CVE-2025-14095.
| URL | Resource |
|---|---|
| https://www.radiometer.com/myradiometer |
|