Description

The ventilator's microcontroller lacks memory protection. An attacker could connect to the internal JTAG interface and read or write to flash memory using an off-the-shelf debugging tool, which could disrupt the function of the device and/or cause unauthorized information disclosure.

INFO

Published Date :

2024-11-14T21:31:14.701Z

Last Modified :

2024-11-18T15:23:48.292Z

Source :

Baxter
AFFECTED PRODUCTS

The following products are affected by CVE-2024-48970 vulnerability.

Vendors Products
Baxter
  • Life2000 Ventilator Firmware